Smadi, Othman (2008) Blood flow dynamics through a defective mechanical heart valve. Masters thesis, Concordia University.
Preview |
Text (application/pdf)
1MBMR40924.pdf - Accepted Version |
Abstract
A model of asymmetrical 2-D defective bileaflet mechanical heart valve was used to simulate steady and pulsatile blood flow through a defective mechanical heart valve under several flow and malfunction severity conditions. The calculations used Reynolds-averaged Navier-Stokes equations with Newtonian blood properties. The results showed that the flow upstream and downstream of the defective valve is highly influenced by malfunction severity and this resulted in a misleading improvement in the correlation between simulated Doppler echocardiographic and catheter transvalvular pressure gradients. In this study, two potential non-invasive parameters were proposed using Doppler echocardiography and phase contrast magnetic resonance imaging, for an early detection of mechanical heart valve malfunction. Finally, the relation between the coherent structures downstream of the valve and the valve malfunction was shown and the significant impact of valve malfunction on platelet activation and as a consequence, on thrombus formation was also shown
Divisions: | Concordia University > Gina Cody School of Engineering and Computer Science > Mechanical and Industrial Engineering |
---|---|
Item Type: | Thesis (Masters) |
Authors: | Smadi, Othman |
Pagination: | xvi, 109 leaves : ill. ; 29 cm. |
Institution: | Concordia University |
Degree Name: | M.A. Sc. |
Program: | Mechanical and Industrial Engineering |
Date: | 2008 |
Thesis Supervisor(s): | Kadem, Lyes |
Identification Number: | LE 3 C66M43M 2008 S63 |
ID Code: | 975743 |
Deposited By: | Concordia University Library |
Deposited On: | 22 Jan 2013 16:14 |
Last Modified: | 13 Jul 2020 20:08 |
Related URLs: |
Repository Staff Only: item control page