Abstract

Freehand three-dimensional ultrasound (3D US) imaging is commonly used for clinical diagnosis and therapy monitoring. In this technique, accurate tracking of the US transducer is a crucial requirement to develop high-quality 3D US volumes. However, current methods for transducer tracking are generally expensive and inconvenient. This thesis presents a low-cost camera-based system for tracking the US transducer with six degrees of freedom (DoF). In this system, two orthogonal cameras with non-overlapped views are mounted on the US transducer. During US
scanning, the two cameras are employed to track artificial features attached to the skin of the patient. A 3D surface map is constructed based on the tracked features and the 3D poses of each camera with respect to the skin are extracted separately. The estimated poses of the two cameras are spatially combined to provide accurate and robust pose estimation of the US transducer. In particular, the fusion of the estimated poses by the two cameras is performed using Kalman filtering based technique,
which is a popular optimization technique in motion guidance and tracking. The camera-based tracking of the US transducer has been applied to synthesize freehand 3D US volumes. The performance of the proposed system is evaluated by performing in-vitro 3D US imaging experiments and quantifying the synthesized US volumes. The results demonstrate that two points in the 3D US volume separated by a distance of 10 mm can be reconstructed with an average error of 0.35 mm and a 3D volume of a cylinder can be estimated within an error of 3.8%.

Divisions:Concordia University > Gina Cody School of Engineering and Computer Science > Electrical and Computer Engineering
Item Type:Thesis (Masters)
Authors:Baba, Mohammad
Institution:Concordia University
Degree Name:M.A. Sc.
Program:Electrical and Computer Engineering
Date:April 2016
Thesis Supervisor(s):Ait Mohamed, Otmane
Keywords:3D Freehand Ultrasound Imaging, Camera-based tracking
ID Code:981081
Deposited By: Mohammad Mustafa Baba
Deposited On:15 Jun 2016 14:31
Last Modified:18 Jan 2018 17:52
All items in Spectrum are protected by copyright, with all rights reserved. The use of items is governed by Spectrum's terms of access.

Repository Staff Only: item control page

Downloads per month over past year

Research related to the current document (at the CORE website)
- Research related to the current document (at the CORE website)
Back to top Back to top